KB024 是否可以在发热发红光的表面上进行测量?

Basically a measurement on hot or glowing surfaces is possible. This has already been implemented in numerous projects using the Dimetix sensors (see Application Example Steel). Depending on the application, additional measures may be required.

It is important to protect the sensor against over temperature. The sensor should therefore be installed at an adequate distance and / or inside a cooled housing.

Hot surfaces emit light over the entire spectral range, and thus also within the red wavelength of the laser (type. 650nm). The higher the temperature is, the higher is also this disturbance. (See picture on the left)

The filter glass integrated in the laser sensors already reduces disturbances outside of the lasers wavelength. In order to even further reduce disturbances, an additional bandpass filter glass can be installed in front of the sensor optics.

For the alignment and selection of the filter glass, follow the instructions Measuring through glass and Minimizing optical disturbances.