Process laser point alignment (Especially for positioning applications):
- Alignment at close range (about 5cm): Align the sensor to the center of the measurement target (horizontal and vertical). Attention: Adjust the position of the sensor / measuring target only in this step. In the following steps only fine adjustment of the sensor.
- Alignment first distance (about 10m): Readjust the laser point to the center of the measurement target. Use the alignment screws of the sensor or the alignment possibilities of the sensor mounting.
- Alignment further larger distances (about 20 / 50 / 100m): Readjust the laser point to the center of the measurement target. Use the alignment screws of the sensor or the alignment possibilities of the sensor mounting.
Note: Typically, alignment is relatively good from about 50m. - Check alignment: Check if the laser spot is on the measurement target for the entire measurement range. If not, restart with step 2.
Note: Film sequence for alignment with Dimetix Sensor, see Dimetix Youtube Channel.